APPLICATION GUIDE
            2.0 MM HS HEADED SCREW AND 2.5 MM HCS COMPRESSION SCREW






               APPLICATION GUIDE
               2.0 MM HS HEADED SCREW AND 2.5 MM HCS COMPRESSION SCREW
               WHEN USING THE SYSTEM ALWAYS PAY ATTENTION TO THE CORRECT COLOR ASSIGNMENT OF THE
               INSTRUMENTS AND SCREWS!

                       2.0 MM HEADED SCREW                     2.5 MM HCS COMPRESSION SCREW
                       Color code: Turquoise blue              Color code: Turquoise blue


                                                                      Kirschner w. trocar point, rd. end, Ø 0.9 mm,
                                                                      L. 80 mm
                                                                      Ref. No. 91-004-0330-009-080
            Drill the Kirschner wire to the desired depth into the bone using an image converter to check the placement.


                                                                      HBS-Measuring Sleeve for Guide Wire
                                                                      Ref. No. 91-INS-8203-00

            Slide the measuring sleeve over the round end of the Kirschner wire, making sure that the tip of the
            instrument rests against the bone. The length of the screw can be determined by means of the scaling on the
            measuring sleeve. Check the length at the height of the end of the round Kirschner wire.


                                                                      HBS-AO cann. drill bit Ø 1.7/1.0 mm, 70 mm,
                                                                      Color code: Turquoise blue
                                                                      Ref. No. 91-INS-8190-17
            Using an image converter to check placement, drill the cannulated drill into the bone over the Kirschner wire.
            We recommend inserting the drill into the bone just short of the tip of the Kirschner wire. If you over-drill
            the tip of the Kirschner wire, it may lose its hold and be removed when the drill is pulled out!



                                                                      Countersink, AO, cann., Ø 4.0/1.1 mm, 60 mm
                                                                      Color code: Turquoise blue / luminous red
                                                                      Ref. No. 91-INS-8192-40
            2.0 HS headed screw
            Optionally it is possible to countersink the screw head into the cortex. Place the cannulated countersink over the
            Kirschner wire and onto the cortex and slowly rotate the countersink into the cortex.
            Important: For osteoporotic bone conditions, it is recommended to perform this step without a power tool. To do this,
            clamp the countersink into the screwdriver handle


                                                                      HBS-AO cann. drill bit Ø 2.5/1.9 mm, WL 4/9 mm
                                                                      extra short, Color code: Turquoise blue
                                                                      Ref. No. 91-INS-8302-04
            2.5 mm HCS compression screw
            Optionally it is possible to deepen the cortex to facilitate screwing in of the head thread of the HCS screws. Guide the
            cannulated extra short HBS drill over the Kirschner wire and slowly rotate the countersink into the cortex.



                                                                      HBS cann. screw driver shaft,
                                                                      HEX 1.5 mm/1.1 mm, 97 mm, AO-conn.,
                                                                      Color code: Turquoise blue/luminous red
                                                                      Ref. No. 91-INS-8304-01-SW1.5
            Connect the corresponding screwdriver shaft with the screwdriver handle. Place the screw on the screwdriver shaft.
            Insert the screw together with the screwdriver over the Kirschner wire and turn the screw clockwise into the bone. Make
            sure the screw comes to rest correctly and firmly on the cortex. Check the correct position of the screw with the help of
            the image converter and then remove the Kirschner wire.