Page 7 - Gipssäge_Autopsie_Absaugung
P. 7
AUTOPSY SAW EPA - ELECTRONIC POWER
AUTOPSY SAW EPA - ELECTRONIC POWER
TECHNICAL DETAILS:
Set Art.-Nr. 30-240-00 Set Art.-Nr. 30-245-00
1 x Autopsy Saw EPA, 230V,50Hz,500W (Art.-No. 30-210-01) 1 x Autopsy Saw EPA, 120V,60Hz,500W (Art.-No. 30-215-01)
1 x Sawblade Ø50 mm for autopsy (Art.-No. 30-260-10) 1 x Sawblade Ø50 mm for autopsy (Art.-No. 30-260-10)
1 x Sawblade Ø65 mm for autopsy (Art.-No. 30-260-13) 1 x Sawblade Ø65 mm for autopsy (Art.-No. 30-260-13)
1 x Sawblade Segment 50/65 mm for autopsy (Art.-No. 30-260-30) 1 x Sawblade Segment 50/65 mm for autopsy (Art.-No. 30-260-30)
1 Paar Wrenches for sawblade changing (Art.-No. 30-198-00) 1 Paar Wrenches for sawblade changing (Art.-No. 30-198-00)
1 x Hexagon key with T-grip (Art.-No. 30-199-00) 1 x Hexagon key with T-grip (Art.-No. 30-199-00)
1 x Grease (filled in a syringe) (Art.-No. 30-200-02) 1 x Grease (filled in a syringe) (Art.-No. 30-200-02)
230 V~, 50 Hz, 500 W 120 V~, 60 Hz, 500 W
3500 – 24000 oscillations, infinitely variable 3500 – 24000 oscillations, infinitely variable
5 meter cable with Euro-plug 3 meter cable with US-plug
• 500 Watt power
• 500 bis 24000 oscillations, infinitely variable
• Fullautomatically power-compensation above the oscillations range
• Slim motor edition with silent motor ca. 65 – 85 dB(A) ± 5 dB(A) (Level 1 to 3) in relation to the selected oscillations,
under unfavourable operating conditions ca. 95 dB(A) ± 5 dB(A) (Level 4 to 6)
• Latest electronic system with various protection and security functions
• Special positioning of the motor
• Robust construction for highest loading capacity by continuous operation
• Safety ON / OFF switch
• Weight only 1,4 kg without cable (1,6 kg with cable)
• Vibrations- and maintenance-free
• Regulation from outside of the manual operating area
• 230 Volt version with 5 meter cable with EU-plug or 110 Volt version with 3 meter cable with US-plug
• Double protection isolation according to class II, medical use of typ BF
• CE conformity
• Tested and conform with IEC CISPR 11:2005 IEC 60601-1:2005+AMD1:2012 IEC 60601-1-2:2014 IEC 60601-1-
6:2010+AMD1:2013 IEC 61000-3-2:2014 IEC 61000-3-3:2013+AMD1:2017 IEC 61000-4-28:1999+AMD1:2001+A
MD2:2009
• National differences EU Group differences EU Special National Conditions EU A-Deviations for CA (Canada) and US
(United Staates of America)
Digimed Medizintechnik • Kreutzerstrasse 1 • 78573 Wurmlingen • Germany • www.digi-med.de 7