Page 31 - Gipssäge_Autopsie_Absaugung
P. 31

AUTOPSY SAW EPA / EPA3K - INSTRUCTION MANUAL




              AUTOPSY SAW EPA / EPA3K ,,ELECTRONIC POWER‘‘
                INSTRUCTION MANUAL

              The load compensation is permanently active and regulates the oscillations, even if various loads are occurring.
              The Speedo on the drive shaft registers a reduction of the oscillations, if the burden of the motor is increased.
              By comparing the selected power and the number of required oscillations, the operation of the system will be adjusted
              permanently. In case of a reduced number of oscillations due to an excess load, the electronic system will re-adjust the
              frequency of oscillations.

              This procedure is done so quickly, that the user won’t recognize any difference. Therefore the plaster saw offers an
              optimised number of oscillations during use. In case of excessive use of the motor (e.g. if saw blades are stuck) the
              electronic system switches off the saw. The saw can be re-used after a short break.

              By using the latest electronic system and electrical parts we are able to offer you a slim motor with a user friendl y
              housing. Furthermore ergonomic and anthroposophic aspects have been considered in respect of smaller hands,
              noiseless during use (as not to frighten children while cutting the plaster, we recommend a lower number of oscillations).
              The ambiental temperature for storing and transportation should not fall below 5 °C and not exceed 40°C. Furthermore
              humidity should not exceed 80%.

              TECHNICAL DATA / CLASSIFICATION
              Part.-Nr. 30-210-01, autopsy saw, standard = 230 V, 50 Hz, 500 W
              Part.-Nr. 30-215-01, autopsy saw, standard = 110 V, 60 Hz, 500 W
              Part.-Nr. 30-220-01, autopsy saw, special triangle chuck = 230 V, 50 Hz, 500 W
              Part.-Nr. 30-225-01, autopsy saw, special triangle chuck = 110 V, 60 Hz, 500 W
              The equipment is in accordance with classe II and typ BF.

              TECHNICAL SPECIFICATION
              •   230 V, 50 Hz, 500 W or
              •   110 V, 60 Hz, 500 W
              •   infinitely variable from 6500 up to 24000 oscillations
              •   regulation from outside of the manual operating area
              •   automatic load compensation for the whole oscillations amplitude
              •   latest electronic system with various protection and security functions
              •   5 m cable with EU-connection, 230 V
              •   3 m cable with US-connection, 110 V
              •   special positioning of the motor, protected by plaster dust
              •   reinforced armature shaft
              •   slim motor edition with silent motor 52 – 77 dB(A) in relation to the selected number of oscillations

              SYMBOL
                   double protection isolation according to class II

                   medical use of typ BF

                   ~ alternating current
                   CE conformity

                   0/1 Power switch Off / On


























            Digimed Medizintechnik • Kreutzerstrasse 1 • 78573 Wurmlingen • Germany • www.digi-med.de      31
   26   27   28   29   30   31   32   33   34   35   36