Page 436 - Digimed_Kataloge_komplett
P. 436

REMOVAL OF ThE „Y“-TYpE LIGhT TROchANTERIc NAILING


















              25  REMOVING OF ThE INTERLOcKING ScREW


                  Remove all the interlocking screws with the

                  screwdriver (22).
















              26  ASSEMBLING OF ThE REMOVAL hITTING DEVIcE



                  Put the slide hammer (8) over the hammer guide shaft
                  (12), close its end with the hammer guide head (9) and
                  secure it with the 12 mm wrench (29).
















              27  reMoVing oF The TrochanTeric nail



                  Put the removal device fixation sleeve (23) over the
                  cardan rod (20) and assemble with the hammer shaft.
                  The implantat can be removed with the slide hammer.

















            Digimed Medizintechnik • Kreutzerstrasse 1 • 78573 Wurmlingen • Germany • www.digi-med.de      21
   431   432   433   434   435   436   437   438   439   440   441