SURGIcAL TEchNIQUE FOR „Y“-TYpE LIGhT TROchANTERIc NAILING










             1   PreoPeraTiVe Planning


                 Preoperative X-ray of the uninjured distal femur is used to
                 estimate proper nail diameter, nail length, and CCD-angle
                 (caput-collumdiaphyseal angle).


                 Fluoroscopic control is advised.
                 The image intensifier must be in a standard position for ap
                 and lateral view.






              !  Numbers in brackets after instrument names refer to the list
                 number in the instrument tray.

                 Pay special attention, important step.









             2   ASSEMBLING OF ThE INSTRUMENTS

                 Attach the intramedullary nail to the radiolucent aiming arm (1) with
                 the nail adapter screw (6). Secure the nail with the 10 mm hex-
                 agonal screwdriver (16, 18). Depending on the CCD angle of the
                 implant (125°/130°/135°) mount the correct aiming arm
                 attachment (2, 3,4) with the aiming arm fixing screw (5).


            !    Drive the fixing screw into the nail, but the tip of the screw
                 should not hinder the rotation of the dynamic screw.







             3   Put into the dynamic screw

                 hole the yellow soft tissue protector (24) and reamer (33).
                 Put into the distal hole the green soft tissue protector (26),
                 drill sleeve (27) and 4 mm spiral drill (28).
















            14          Digimed Medizintechnik • Kreutzerstrasse 1 • 78573 Wurmlingen • Germany • www.digi-med.de